During the Heart Rhythm Scientific Sessions this week, Boston Scientific announced positive LBCT data underscoring the safety and efficacy of multi-site pacing (MSP) over conventional cardiac resynchronization therapy (CRT), as well as poster presentations on real-world use of the WATCHMAN Left Atrial Appendage Closure (LAAC) device and next-generation WATCHMAN FLX LAAC device as alternatives to oral anticoagulation therapy for stroke risk reduction in patients with non-valvular atrial fibrillation.

What are the Most Commonly Performed Surgeries?

Whether you are trying to learn more about a surgery you are planning to have or find out what the most common procedures are, this list is here for you. Read more.

B2B Training: What’s the Most Effective Way of Learning?

B2B training, which is short for business-to-business training, is a popular and convenient type of training that enables entire companies to train their teams...

Trividia Health, Inc. today announced it is initiating a nationwide voluntary recall of one (1) TRUE METRIX® AIR Blood Glucose Meter distributed in the United States to one customer (AssuraMed). The meter was not packaged into store-brand or retail branded packaging.

The company has determined that one (1) isolated TRUE METRIX AIR Blood Glucose Meter was packaged into a TRUE METRIX Blood Glucose Meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL. If a consumer does not notice the incorrect unit of measure, it is possible that the meter glucose result will be read as a lower blood glucose result than expected, and this may result in the patient’s glucose level remaining high, which can lead to serious injury or impairment with risk of death.

There is one (1) affected TRUE METRIX AIR meter with serial number TA1548753 that was packaged into a TRUE METRIX Blood Glucose Meter Kit with the Lot number KW0135 and with UDI # (01)00021292006075(17)200831(10)KW0135(21)TA1548753. It has been determined that this kit was distributed nationwide in the United States in February 2019. Trividia Health has not received any reports of patient injury or an adverse event related to this voluntary recall.

Consumers who have or may have the TRUE METRIX AIR meter should do the following:

  • Check to confirm if you have the affected TRUE METRIX AIR meter by obtaining the meter serial number (TA1548753) from the serial number label located on the back of the meter, and then visit www.TrividiaHealth.com/air-product-notice or call Trividia Health toll-free at 1-800-518-5726, Monday-Friday, 8AM-8PM EST.
  • If you have the affected TRUE METRIX AIR meter, please stop using the meter and call Trividia Health immediately to verify the serial number, and to expedite return and replacement of the affected TRUE METRIX AIR meter at no charge.
  • Consumers may continue to test blood glucose using any other Trividia Health blood glucose meter not included in this voluntary recall while waiting for their replacement meter to arrive. Use only the test strips that are intended for use with your blood glucose meter.

Notifications will be sent to US pharmacies, durable medical equipment providers and distributors who may have received this device from Trividia’s customer.

Patient safety is a top priority at Trividia Health. The company has notified the U.S. Food and Drug Administration (FDA) of this voluntary product recall and is working with the customer who received the affected meter to quickly resolve this matter.

Users within the USA may report adverse reactions or quality problems experienced with the use of this product to the FDA’s MedWatch Adverse Event Reporting program either online or by phone.

spot_img

DON'T MISS

Mediclinic Appoints 2 New Independent Non-Executive Directors

Natalia Barsegiyan and Zarina Bassa will bring significant and diverse skills and experience, complementing those of existing Board members. I am confident they will both be great additions to our Board of Directors.”

Asensus Surgical Appoints Two New Board Members

Asensus Surgical, Inc. (NYSE American: ASXC), a medical device company that is digitizing the interface between the surgeon and the patient to pioneer a new era of Performance-Guided Surgery™, today announced the expansion of its Board of Directors, appointing...

Our Sister Publication: Biotechnology News Magazine

Subscribe to Medical Device News Magazine here.