Philips Provides Update on their Respironics’ PE-PUR Sound abatement Foam Test & Research Program

Testing in accordance with ISO 18562 and ISO 10993 standards conducted by five certified, independent testing laboratories in the US and Europe

On June 14, 2021, Royal Philips’ (NYSE: PHG; AEX: PHIA) subsidiary Philips Respironics, initiated a voluntary recall notification/field safety notice* to address potential health risks related to the polyester-based polyurethane (PE-PUR) sound abatement foam in specific CPAP, BiPAP and mechanical ventilator devices.

At the time the recall notification/field safety notice* was issued, Philips Respironics relied on an initial, limited data set and toxicological risk assessment, and assumed a worst-case scenario for the possible health risks out of an abundance of caution. Since then, together with five certified, independent testing laboratories in the US and Europe, as well as other qualified third-party experts, Philips Respironics has been conducting a comprehensive test and research program on the PE-PUR foam to better assess and scope the potential patient health risks related to possible emission of particulates from degraded foam and volatile organic compounds (VOCs).

This update is intended to provide healthcare providers, patients, and other stakeholders with updated information on the testing results to date. Philips will continue to provide regular updates as new test results and assessments become available, as not all tests have been completed to date.

The overall guidance for healthcare providers and patients in the recall notification/field safety notice* remains unchanged at this time.

Testing methods

Testing results and conclusions to date are grouped by device air path design and configuration, i.e., based on how the air flows through the device. Of the five device categories, the first-generation DreamStation devices represent 68% of the registered affected devices globally. Within each device category, testing and analyses are performed on new devices with pristine foam, devices with lab-aged foam and used devices:

  • Visual assessment of the foam in returned/used devices to assess the prevalence of visible foam degradation.
  • VOC testing to identify and quantify organic compounds that may be inhaled during device use.
  • Particulate Matter (PM) testing to determine concentrations of respirable particulates, i.e., particulates up to 10 micrometers in diameter, as it relates to inhalation risks and established health thresholds.
  • Additional physical, chemical and biological testing of the PE-PUR foam related to patient risks if patients were in contact with foam material.

The complete update on the PE-PUR testing results and conclusions available to date can be found here, and the main findings have been presented below. Healthcare providers, patients, and other stakeholders should use the complete update for any informed decision making, and not the overview in this press release.

“I deeply regret the concern experienced by patients who rely on the affected Respironics sleep and respiratory devices for their health and quality of life and want to emphasize our commitment to providing them with a resolution as fast as possible,” said Frans van Houten, CEO of Royal Philips. “More than 1,000 of our colleagues are working extremely hard to achieve this. While certain prolonged tests across the affected product categories are still to be completed, the results to date for the first-generation DreamStation devices, which represent the majority of the registered affected devices, show a very low prevalence of visible foam degradation. In addition, the new and used first-generation DreamStation devices passed volatile organic compound and respirable particulate emission testing. This is very encouraging. Results to date also indicate that ozone cleaning significantly exacerbates foam degradation.”

First-generation DreamStation devices (68% of registered affected devices globally)

Visual assessment: In order to determine the prevalence of foam degradation, which may result in foam particulate emission, a visual assessment of the foam was performed on a sample of 60,847 returned/used first-generation DreamStation devices from the US and Canada. The visual inspection was conducted according to a specific protocol as part of the repair process. The sample included 36,341 devices for which the users reported no use of ozone cleaning, 11,309 devices for which the users reported use of ozone cleaning, and 13,197 devices for which it was reported unknown by the user whether ozone cleaning was used.

  • Prevalence of visible foam degradation (US and CA): 164 of 36,341 (0.5%) devices with self-reported no ozone use showed significant visible foam degradation.
  • Impact of repeated ozone cleaning: Devices with self-reported ozone use were 14x more likely to have significant visible foam degradation than those with self-reported no ozone use: 777 of 11,309 devices (7%) showed significant visible foam degradation.
  • Assessment of a sample of devices linked to reported alleged foam degradation complaints: 422 devices of the inspected 60,847 returned/used devices are linked to a reported foam degradation complaint. However, only 18 out of these 422 devices (4%) actually showed visible foam degradation.

In those devices where visible foam degradation was significant, i.e., there was reduction in foam volume, it was observed that there was accumulation of degraded foam within the airpath inside the device. The foam becomes hygroscopic (i.e., absorbs moisture) and sticky with degradation. It also loses significant volume and increases density as the structure changes from a foam to a viscous liquid material. As such, even when foam particulates are formed by degradation, they are likely to accumulate within the device and may not be directly emitted by the device.

A visual assessment of the foam was also performed on a sample of 1,360 returned/used first-generation DreamStation devices from various countries in Europe and on a sample of 931 returned/used devices from Japan. 

  • Prevalence of visible foam degradation (EU and JPN): None of the assessed devices from Europe or Japan showed significant visible degradation.

Volatile Organic Compounds (VOCs): VOC testing according to ISO 18562-3 was performed on new, lab-aged and used devices to (1) quantify VOC emissions from devices, and (2) assess the toxicological risk associated with exposure to the quantified concentrations of those VOCs. It is important to note that these tested new and lab aged first-generation DreamStation devices were not exposed to ozone cleaning, in accordance with the instructions for use.

  • As previously provided in an update on December 23, 2021, VOC emissions are below established limits based on ISO 18562-3 testing and evaluation of new, lab-aged, and used devices. Exposure to the level of VOCs identified to date for the first-generation DreamStation devices is not anticipated to result in long-term health consequences for patients.

Particulate Matter (PM): PM testing according to ISO 18562-2 was performed on the devices to (1) quantify the particulate matter emitted from devices, and (2) assess whether the concentration detected is less than thresholds provided in the standard.

  • New devices and used devices (including used devices with visible foam degradation) were tested and were all found to be compliant with ISO 18562-2 allowable limits for PM emissions
  • Tested PM emissions of used devices with degradation were not statistically different than PM emissions without degradation, suggesting that degradation did not contribute to appreciable elevated levels of respirable particles in the devices tested
  • The used devices that were tested for PM emissions, were also evaluated for cleanliness based on a visual inspection of the exterior of the device, i.e., the presence of environmental materials on the external surface of the device, such as the inlet filter location. For these devices, average particulate matter counts in devices classified as ‘dirty’ were significantly greater than those classified as ‘clean’.

Biocompatibility testing of (degraded) PE-PUR foam: Additional testing is still being performed in accordance with ISO 10993 to facilitate a toxicological risk assessment of (degraded) foam particulates, which is relevant if they would potentially contact the patient. This testing includes chemical characterization (i.e., what chemicals may potentially extract or leach from the foam and have direct contact with body tissues and/or fluids), in vitro assessment (i.e., tests performed in a test tube, dish, etc. outside the body), and in vivo assessment (i.e., preclinical testing) of new, lab aged and/or used PE-PUR foam.

To support the assessment of potential genotoxicity, cytotoxicity, and irritation risks of lab-aged foam, chemical characterization of lab-aged foam, as well as experiments to assess the amount of foam that can potentially contact the patient are being conducted, as lab-aged foam did not pass Ames (genotoxicity), cytotoxicity and skin irritation bioassay testing. Further, complementing the lab-aged foam assessment, a chemical characterization of returned/used degraded foam is also being conducted to better elucidate risks under field conditions. Per ISO 10993, the bioassay results cannot stand alone and, therefore, a positive Ames, cytotoxicity, or skin irritation result triggers a required follow-up evaluation including identification of potential confounding factors, and a weight of evidence assessment to determine a confirmed conclusion on potential risks for patients under expected usage of the device.

Other devices under the recall notification/field safety notice

Other devices that are being tested include DreamStation Go (1% of the registered devices) and SystemOne (26% of the registered devices). These devices each have a different air path design/configuration compared to the first-generation DreamStation devices but contain the same PE-PUR foam.

New DreamStation Go and SystemOne devices passed VOC and PM testing based on standards available prior to ISO 18562 i.e., Indoor Air Quality Evaluation (as previously disclosed in the April 25, 2022, update). Further ISO 18562 VOC and PM testing is ongoing. The results of ISO 10993 biocompatibility testing of degraded PE-PUR foam for the first-generation DreamStation devices, which is ongoing, will also apply to DreamStation Go and SystemOne devices.

Further devices include Trilogy 100/200 (3% of the registered devices) and OmniLab/A-Series BiPAP (2% of the registered devices). New Trilogy 100/200 passed VOC and PM testing. New OmniLab devices passed VOC and PM testing based on standards available prior to ISO 18562 i.e., Indoor Air Quality Evaluation (as previously disclosed in the April 25, 2022, update). New and used OmniLab devices passed VOC testing based on ISO 18562. Further testing is ongoing.

Summary of ongoing tests

The first-generation DreamStation, DreamStation Go and SystemOne CPAP/BiPAP devices represent 95% of the registered affected devices globally. Philips Respironics expects to complete the remaining VOC and PM testing for these devices, as well as the degraded foam toxicological risk assessments (in accordance with ISO 10993) in the coming months. Philips Respironics will also continue with the tests to assess the impact of repeated ozone cleaning on foam degradation in these CPAP/BiPAP devices, as well as the remaining VOC and PM testing and the degraded foam toxicological risk assessments for the Trilogy 100/200 and OmniLab ventilator devices.

Silicone foam testing

In November 2021, the FDA requested that Philips retain an independent laboratory to perform additional testing to determine what, if any, potential safety risks may be posed to patients by silicone-based foam. Philips Respironics engaged independent testing laboratories to perform additional VOC testing. Based on the draft reports, Philips Respironics has not identified any safety issues. The assessment is being completed, and the final reports are subject to FDA review, which are expected in the coming months.

Guidance for healthcare providers and patients

As indicated, this update is intended to provide healthcare providers, patients, and other stakeholders with updated information on the testing results to date. The overall guidance for healthcare providers and patients in the recall notification/field safety notice* remains unchanged at this time. Philips Respironics remains fully committed to addressing all devices affected by the recall notification/field safety notice* and continues to work with the relevant competent authorities to further optimize the remediation plan. To date, approximately 2.7 million replacement devices and repair kits have been produced.

Additional information

Further information, including the complete update and FAQs, as well as video messages from Philips CEO Frans van Houten, Chief Business Leader Connected Care Roy Jakobs and Technical Project Manager for the test and research program Jan Bennik, can be found here.

Hot this week

Cartessa Aesthetics Partners with Classys to Bring EVERESSE to the U.S. Market

Classys, which is listed on the KOSDAQ, is one of South Korea's most distinguished aesthetic technology manufacturers, with devices distributed in 80+ markets globally. This partnership marks Classys's official entry into the American marketplace, with Cartessa Aesthetics as the exclusive distributor for EVERESSE, launched under the Volnewmer brand in current global markets.

Stryker Launches Next-Generation of SurgiCount+

Now integrated with Stryker's Triton technology, SurgiCount+ addresses two key challenges: retained surgical sponges and blood loss assessment. Integrating these previously separate digital solutions provides the added benefit of a more efficient, streamlined workflow for hospitals notes Stryker.

Nevro Receives CE Mark In Europe for It’s HFX iQ™ Spinal Cord Stimulation System

Nevro notes HFX iQ is the first and only SCS system with artificial intelligence (AI) technology that combines high-frequency (10 kHz) therapy built on landmark evidence that uses ongoing cloud data insights to deliver personalized pain relief

Recor Medical Reports: CMS Grants Distinct TPT Device Code and Category to Recor Medical for Ultrasound Renal Denervation

The approval of TPT offers incremental reimbursement payments for outpatient procedures performed with ultrasound renal denervation for Medicare fee-for-service beneficiaries. It becomes effective January 1, 2025, and is expected to remain effective for up to three years notes Recor Medical.

Jupiter Endovascular Reports | 1st U.S. Patient Treated with Jupiter Shape-shifting Thrombectomy Device

“Navigation challenges during endovascular procedures are often underappreciated and have led to under-adoption of life-saving procedures, such as pulmonary embolectomy. We have purpose-built our Endoportal Control technology to solve these issues and make important endovascular procedures accessible to more clinicians and their patients who can benefit from them,” said Carl J. St. Bernard, Jupiter Endovascular CEO. “This first case in the U.S. could not have gone better, and appears to validate the safety and performance we are seeing in our currently-enrolling European SPIRARE I study.”