Smiths Medical Issues Worldwide Notification Regarding Potential for Inappropriate Bolus or Loading Dose Delivery in Certain Medfusion® 3500 and 4000 Infusion Pumps

Smiths Medical has become aware of an issue where certain Medfusion® 3500 and 4000 syringe pumps with firmware versions listed below may not operate as expected.

Smiths Medical reports due to a software error, if a bolus or loading dose is interrupted and a specific sequence of events occurs, over- or under- delivery of a bolus or loading dose may occur.

Affected Models:

Medfusion® 3500 Syringe Pump – Firmware Version 6.0.0
Medfusion® 4000 Syringe Pump – Firmware Versions 1.5.0; 1.5.1; 1.6.0; 1.6.1; 1.7.0

Affected Models listed above were distributed by Smiths Medical from 2013 to 2020. The firmware version of these pumps may be identified by powering them on and observing the firmware version displayed on the screen. This communication, and the associated Recall Notice issued by Smiths Medical to consignees of Affected Models, is applicable only to the Affected Models listed above.

In the Affected Models listed above, when an operator: (1) Programs a Bolus or Loading Dose, (2) Initiates the Bolus or Loading Dose, (3) Interrupts the Bolus or Loading Dose, (4) Primes the pump, and (5) Resumes the Bolus or Loading Dose instead of beginning a new infusion, over- delivery or under-delivery of medication may occur.

WARNING: Continuing an interrupted bolus or loading dose may result in medication over- delivery or under-delivery which can result in serious injury or death.

INSTRUCTIONS TO CLINICIANS

For any reason, when a bolus or loading dose delivery is interrupted and mechanically primed using the PRIME FUNCTION, cancel the bolus or loading dose and begin a new infusion to avoid potential over-delivery or under-delivery of medication.

This action has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA). Smiths Medical has received a total of 4 reports associated with this issue, all involving over-delivery of medication. One of these reports resulted in a serious injury; there were no reports of deaths.

Smiths Medical is issuing Recall Notices and Response Forms to consignees in possession of Affected Models and will notify consignees of Affected Models when software updates are available to correct this issue. Software updates will be provided at no cost to all consignees and are expected to be available in late 2021.

FDA MedWatch Reporting

Adverse reactions or quality problems experience with the use of these products may also be reported to the FDA’s MedWatch Program by:

  • MedWatch website at www.fda.gov/medwatch
  • Phone: 1-800-FDA-1033
  • Fax: 1-800-FDA-0178
  • Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD20852

Hot this week

Avery Dennison Medical Introduces Ipdated SilFoam Lite: Sustainability, MDR Certification & Performance Improvements

The newly enhanced SilFoam Lite delivers superior efficiency and reliability, bringing improved fluid handling capabilities and improved tack. These improvements make the product ideal for customers seeking quality, high-performance solutions in wound care notes Avery Dennison Medical.

Voluntary Recall Notifying Medtronic Insulin Pump Users of Potential Risks of Shortened Pump Battery Life

Medtronic plc voluntarily issued a field action starting on July 31, 2024, notifying global customers of its MiniMed™ 600 series or 700 series insulin pumps to follow their pump's built-in alerts and alarms for battery status and to contact Medtronic if they observe changes in the battery life of their pump

Medtronic Expands AiBLE Spine Surgery Ecosystem with New Technologies and Siemens Healthineers Partnership

New advancements in the AiBLE Spine Surgery ecosystem build upon the company's commitment to procedural innovation and execution

Axlab, Danish Medtech Pioneer, expands to US with Advanced Robotic Tissue Sectioning for Pathology Laboratories

Kris Rokke, National Sales Director for Axlab in the US. "My team and I are extremely excited and honored about this unique opportunity to also offer this advanced technology to labs across the US and thus contribute to the pathology labs of tomorrow."

Spartan Medical Broadens Single-Use Sterile Instrument Portfolio to Improve Outcomes, Increase Efficiency, and Generate Cost Savings

Spartan Medical products portfolio of single-use, sterile med tech includes micro and minor surgical convenience kits, kerrison rongeurs, spinal and general surgical retractors, dural repair kits, synthetic biologics, and a wide range of orthopedic pre-sterilized implants and devices.