No Carolina / NY / Florida
Ph: 561.316.3330

No Carolina | NY | Florida

Smiths Medical Issues Worldwide Notification Regarding Potential for Inappropriate Bolus or Loading Dose Delivery in Certain Medfusion® 3500 and 4000 Infusion Pumps

Editor: What To Know

  • Smiths Medical reports due to a software error, if a bolus or loading dose is interrupted and a specific sequence of events occurs, over- or under- delivery of a bolus or loading dose may occur.
  • (1) Programs a Bolus or Loading Dose, (2) Initiates the Bolus or Loading Dose, (3) Interrupts the Bolus or Loading Dose, (4) Primes the pump, and (5) Resumes the Bolus or Loading Dose instead of beginning a new infusion, over- delivery or under-delivery of medication may occur.
  • For any reason, when a bolus or loading dose delivery is interrupted and mechanically primed using the PRIME FUNCTION, cancel the bolus or loading dose and begin a new infusion to avoid potential over-delivery or under-delivery of medication.

Smiths Medical has become aware of an issue where certain Medfusion® 3500 and 4000 syringe pumps with firmware versions listed below may not operate as expected.

Smiths Medical reports due to a software error, if a bolus or loading dose is interrupted and a specific sequence of events occurs, over- or under- delivery of a bolus or loading dose may occur.

Affected Models:

Medfusion® 3500 Syringe Pump – Firmware Version 6.0.0
Medfusion® 4000 Syringe Pump – Firmware Versions 1.5.0; 1.5.1; 1.6.0; 1.6.1; 1.7.0

Affected Models listed above were distributed by Smiths Medical from 2013 to 2020. The firmware version of these pumps may be identified by powering them on and observing the firmware version displayed on the screen. This communication, and the associated Recall Notice issued by Smiths Medical to consignees of Affected Models, is applicable only to the Affected Models listed above.

In the Affected Models listed above, when an operator: (1) Programs a Bolus or Loading Dose, (2) Initiates the Bolus or Loading Dose, (3) Interrupts the Bolus or Loading Dose, (4) Primes the pump, and (5) Resumes the Bolus or Loading Dose instead of beginning a new infusion, over- delivery or under-delivery of medication may occur.

WARNING: Continuing an interrupted bolus or loading dose may result in medication over- delivery or under-delivery which can result in serious injury or death.

INSTRUCTIONS TO CLINICIANS

For any reason, when a bolus or loading dose delivery is interrupted and mechanically primed using the PRIME FUNCTION, cancel the bolus or loading dose and begin a new infusion to avoid potential over-delivery or under-delivery of medication.

This action has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA). Smiths Medical has received a total of 4 reports associated with this issue, all involving over-delivery of medication. One of these reports resulted in a serious injury; there were no reports of deaths.

Smiths Medical is issuing Recall Notices and Response Forms to consignees in possession of Affected Models and will notify consignees of Affected Models when software updates are available to correct this issue. Software updates will be provided at no cost to all consignees and are expected to be available in late 2021.

FDA MedWatch Reporting

Adverse reactions or quality problems experience with the use of these products may also be reported to the FDA’s MedWatch Program by:

  • MedWatch website at www.fda.gov/medwatch
  • Phone: 1-800-FDA-1033
  • Fax: 1-800-FDA-0178
  • Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD20852
Medical Device News Magazinehttps://infomeddnews.com
Medical Device News Magazine provides breaking medical device / biotechnology news. Our subscribers include medical specialists, device industry executives, investors, and other allied health professionals, as well as patients who are interested in researching various medical devices. We hope you find value in our easy-to-read publication and its overall objectives! Medical Device News Magazine is a division of PTM Healthcare Marketing, Inc. Pauline T. Mayer is the managing editor.

Recent News

Coya Therapeutics Presents Updated ALS Biomarker Data at the 2nd Annual Johnson Center Symposium

Coya Therapeutics, Inc. (Nasdaq: COYA) (“Coya” or the “Company”), a clinical-stage biotechnology company developing biologics intended to enhance regulatory T cell (Treg) function, announces...

How Can Wearable Health Technology Boost Driver Safety and Well-being?

Wearable health technology can measure a host of vital information, including heart rate, heart rate variability, and blood pressure. The information it provides can be a key aid for drivers wishing to stay safe.

VeraClinic: Pioneering Personal Transformation Through Hair Transplantation and Cosmetic Surgery in Turkey

As a testament to Turkey's legacy in the world of aesthetic medicine, VeraClinic invites you to discover the transformative power of our hair transplantation and cosmetic surgery services.
Medical Device News Magazinehttps://infomeddnews.com
Medical Device News Magazine provides breaking medical device / biotechnology news. Our subscribers include medical specialists, device industry executives, investors, and other allied health professionals, as well as patients who are interested in researching various medical devices. We hope you find value in our easy-to-read publication and its overall objectives! Medical Device News Magazine is a division of PTM Healthcare Marketing, Inc. Pauline T. Mayer is the managing editor.

Stay Connected

spot_img

About Medical Device News Magazine

About Medical Device News Magazine. A digital publication founded in 2008 located in the United States. The publication is one of the industry’s leading sources of medical device and biotech industry updates. Medical Device News Magazine is easily accessible 24/7/365 and is a fast 1, 2, 3 easy read! Our purpose is...

By using this website you agree to accept Medical Device News Magazine Privacy Policy